Caring for Your Defibrillator

To discharge and dispose of a used battery charger:

1 Insert the discharger into the used battery

charger.

Note:

Do not attempt to remove the discharger

once it is in place.

2

Let the discharger fully deplete the battery
charger. Wait at least 9 days.

3

Place the used battery charger in the trash or
recycle it.