TRIGGER:

Pins 1 and 2 can be used to apply external trigger sources to the power supply. These pins can

also be used to control the list operation

RI/DFI:

The Inhibit Input pin can be used to control the output state of the power supply (RI function).

The Fault Output pin (DFI function) can be used to indicate internal faults of the power supply.


NOTE
: The Fault indicator function (DFI) is only available for models 9120A, 9121A, 9122A, 9123A, and
9124.

DIGITAL I/O: Read and control output and input state of the 2 available pins. For models 9150, 9151, 9152,

and 9153, only digital input is available.

For Models: 9120A, 9121A, 9122A, 9123A, 9124

Mode
Pin

Trigger

RI/DFI

DIGITAL I/O

1 (INH)

Trigger in

Inhibit Input

Digital Input

2 (GND)

GND

GND

GND

3 (FLT)

Not Used

Fault Indicator Output

Digital Output

4 (GND)

Not Used

GND

GND

For Models: 9150, 9151, 9152, 9153

Mode
Pin

Trigger

RI

DIGITAL I/O

1 (TRIN)

Trigger in

Inhibit Input

Digital Input

2 (GND)

GND

GND

GND

Trigger Operation

The power supply supports 3 different trigger modes. Immediate, External and Bus. Configure one of the trigger

sources before performing trigger operation.

• Trigger Key:

When this function is enabled, you can generate an immediate trigger pulse by pressing

Shift

Trigger.

• External trigger signal (TTL):

When this function is enabled, the power supply can be triggered with a TTL pulse applied to pin 1 of

the terminal connector in the rear. The TTL on pulse width should be at least 5 ms.

• Bus:

When this function is enabled, you can trigger the power supply by sending a *TRG or TRIgger

command to the power supply

Remote Inhibit (RI) - Input

Used to turn off the output of the power supply. Can be used to turn off several power supplies simultaneously.