STEP List operation is step mode.

Command syntax: [SOURce:]LIST:MODE <CRD>

Parameter: CONTinious|STEP

Query syntax: [SOURce:]LIST:MODE?

Return parameter: <CRD>

[SOURce:]LIST:STEP

This command sets the operation mode of the list file.

ONCE List operate once
REPeat Repeat list operation

Command syntax: [SOURce:]LIST:STEP <SRD>

Parameter: ONCE|REPeat

Query syntax: [SOURce:]LIST:STEP?

Return parameter: <CRD>

Note to usage:

When setting to LIST:STEP REPEAT, the list would run repeatedly for an indefinite amount of time.
Below is an example set of commands to setup a repeated list properly:

*IDN?
PORT:FUNCTION TRIGGER
OUTPUT:TIMER OFF
OUPUT OFF
LIST AREA 4
LIST:COUNT 2
LIST:UNIT SECOND
TRIGGER:SOURCE BUS
LIST:VOLTAGE 1,5
LIST:VOLTAGE 2,1
LIST:WIDTH 1,4
LIST:WIDTH 2,2
LIST:MODE CONT
LIST:STEP REPEAT
LIST:NAME ‘MX’
LIST:SAVE 1
MODE LIST
OUTPUT ON
TRIGGER


[SOURce:]LIST:COUNt

This command sets the number of steps for the list operation.

Command syntax: [SOURce:]LIST:COUNt <NRf>

Parameter: 2~400

Query syntax: [SOURce:]LIST:COUNt?

Parameter: None

Return parameter: <NR1>

[SOURce:]LIST :CURRent[:LEVel]

This command sets the current step.

Command syntax: [SOURce:]LIST:CURRent[:LEVel]

Parameter: 0~30A

Unit: A mA

Example: LIST:CURR 1,3A;

Query syntax: [SOURce:]LIST:CURRent:[LEVel]?