B&K Precision 9150-9153 - Manual User Manual

Page 27

Advertising
background image

27





Data Frame Format

Parity=None

Start

Bit

8 Data Bits

Stop

Bit

Parity=Odd,

Even

Start

Bit

8 Data Bits

Parity

Bit

Stop

Bit



End of String character is ’\n’ (0x0a)

DB9 Interface Details

The DB9 connector in the rear panel of the power supply provides a TTL level signal .It can be connected to a

standard PC interface via the IT-E131, IT-E132 or IT-E135 isolated converter/adapter.


Flow Control


There are no hardware flow controls for sending and receiving commands to the power supply.
The average time it takes to both send and receive every command is approximately 200ms. In
the case of more complex commands, more time may be required to complete transmission.

Note: Configuration of address parameter

Communication via RS232 and USB: Setting of the address is optional. It is not required to

communicate with the instrument. The address can be set from the front panel and is stored in non

volatile memory. This feature is useful when communicating via USB, and connecting several

instruments, e.g. via a USB hub. In this scenario, Windows assigns a virtual COM port to each

device which is unknown prior to establishing communications with the instrument (could be

different each time). In this case, the user can correlate each virtual COM port randomly assigned

by Windows with a user defined address.

Communication via GPIB: Setting of the address value is mandatory and corresponds to GPIB

instrument address according to GPIB/IEEE-488 conventions. Valid values are 0 – 30.

Advertising
This manual is related to the following products: