4 output verification, Voltage output check – B&K Precision 9150-9153 - Manual User Manual

Page 8

Advertising
background image

8

The first row displays the actual output voltage value and current and the state of power supply. The second row

displays the voltage measured by the DVM (on left) and the Set Value for the voltage of the power supply.

To obtain additional information about the instrument, press and hold the SHIFT button during Power Up. On

the display you will see the following:

First row:

V/A rating and DVM voltage range

second row: Firmware version and serial number




Press “Esc” to exit the display.

In case the self test routine is not successful, you may see one of the following:

If the EEPROM was damaged or the latest operation data is lost, the VFD will display:

If the calibration data stored in the EEPROM is lost, the VFD will display



If the latest operating state of the power supply in EEPROM is lost, the VFD will display:



Warning: The power supply is shipped from the factory with a power-line cord that has a plug

appropriate for your location. Your power supply is equipped with a 3-wire grounding type power cord; the

third conductor being the ground. The power supply is grounded only when the power-line cord is plugged

into an appropriate receptacle. Do not operate your power supply without adequate cabinet ground

connection.

1.4 Output Verification

The following procedures verify that the power supply outputs the correct voltage and current levels and properly

responds to entries from the front panel.

Voltage Output Check

The following steps verify basic voltage functions without load.

1) Turn on the power supply.

2) Enable the outputs

Press the key. Notice the CV annunciator turning on.

ERR EEPROM

ERROR CAL

Error Config Data

Out on/off

Sourc: XXV XA Meas: XXV
Ver: 1.67 SN:5975002002