Before connecting to an AC outlet or external power source, be sure that the power switch is

in the OFF position and verify that the AC power cord, including the extension line, is

compatible with the rated voltage/current and that there is sufficient circuit capacity for the

power supply. Once verified, connect the cable firmly.

WARNING:

The included AC power cord is safety certified for this

instrument operating in rated range. To change a

cable or add an extension cable, be sure that it can

meet the required power ratings for this instrument.

Any misuse with wrong or unsafe cables will void the

warranty.

2.2 Output Connections

The waveform generator output circuits operate as a 50 Ω voltage source working into a 50

Ω load. At higher frequencies, a non-terminated or improperly terminated output may cause

aberrations on the output waveform. In addition, loads with an impedance less than 50 Ω

will reduce the waveform amplitude, while loads with an impedance greater than 50 Ω will

increase waveform amplitude.

Excessive distortion or aberrations caused by improper termination are less noticeable at

lower frequencies, especially with sine and triangle waveforms. To ensure waveform

integrity, follow these precautions:

1. Use good quality 50 Ω coaxial cable and connectors.

2. Make all connections tight and as short as possible.

3. Use good quality attenuators, if it is necessary to reduce waveform amplitudes

applied to sensitive circuits.

4. Use termination or impedance-matching devices to avoid reflections.

5. Ensure that attenuators and terminations have adequate power handling

capabilities.

If there is a DC voltage across the output load, use a coupling capacitor in series with the

load. The time constant of the coupling capacitor and load must be long enough to maintain

pulse flatness.

Impedance Matching

If the waveform generator is driving a high impedance, such as a 1 MΩ input impedance

(paralleled by a stated capacitance) of an oscilloscope vertical input, connect the

transmission line to a 50 Ω attenuator, a 50 Ω termination and to the oscilloscope input. The

attenuator isolates the input capacitance of the device and terminates the waveform

generator properly.