Check model and firmware version, Output check – B&K Precision 4052-4055 - Manual User Manual

Page 17

Advertising
background image

17

4.

Self Adjust

This option runs an internal self adjustment procedure that will check and adjust the

instrument to specification. Use this option only if the instrument is suspected to be

out of calibration.

Press Utility, select 1/2 from the menu to enter the second menu page, and select

Test/Cal option. Then, select SelfAdjust option. The instrument will run the

procedure internally, and a progress bar will be displayed in the waveform

information area. When finished, it will show 100% on the progress bar and prompt

you to press any function key to continue. Follow the prompt to return to the main

display.

Check Model and Firmware Version

The model and firmware version can be verified from within the menu system.

Press Utility, select 1/2 from the menu to enter the second menu page, and select EditInfo

option. The software/firmware version, hardware version, model, and serial number will be

displayed. Press any function key to exit.

Output Check

Follow the steps below to do a quick check of the settings and waveform output.

1. Turn on the instrument and set the instrument to default settings. To set to default,

press Utility, press 1/2 to go to page 2, select System from the menu, and select Set

to Default. The instrument will set both channels with the following parameters:

Waveform Shape: Sine

Frequency: 1.000000 kHz

Amplitude: 4.000 Vpp

Offset: 0.0 mVdc

Phase: 0.0 °

Output Impedance: Hi-Z