5 specifications, Specifications – B&K Precision 4052-4055 - Manual User Manual

Page 71

Advertising
background image

71

5 Specifications

Note: All specifications apply to the unit after a temperature stabilization time of 15

minutes over an ambient temperature range of 23 °C ± 5 °C. Specifications are subject to

change without notice.

Environmental Conditions:

This instrument is designed for indoor use and operated with maximum relative humidity of

95%.

Model

4052

4053

4054

4055

Channels

2

Frequency Characteristics

Sine

1 µHz – 5 MHz

1 µHz – 10

MHz

1 µHz – 25

MHz

1 µHz – 50

MHz

Square

1 µHz – 5 MHz

1 µHz – 10

MHz

1 µHz – 25 MHz

Triangle, Ramp

1 µHz – 300 kHz

Pulse

500 µHz – 5 MHz

Gaussian Noise (-3 dB) > 5 MHz

> 10 MHz

> 25 MHz

> 50 MHz

Arbitrary

1 µHz – 5 MHz

Accuracy

± 50 ppm (90 days)

± 100 ppm (1 year)

Resolution

1 µHz

Arbitrary Characteristics

Built-in Waveforms

48 Built-in waveforms (includes DC)

Waveform Length

16,000 points / Ch

Vertical Resolution

14 bits

Sampling Rate

125 MSa/s

Minimum Rise/Fall

Time

7 ns (typical)

Jitter (pk-pk)

8 ns (typical)

Non-volatile memory

storage

10 waveforms

Output Characteristics

Amplitude Range

Channel 1: 2 mVpp – 10 Vpp into 50 Ω, ≤ 10 MHz

2 mVpp – 5 Vpp into 50 Ω, > 10 MHz

4 mVpp – 20 Vpp into open circuit, ≤ 10 MHz

4 mVpp – 10 Vpp into open circuit, > 10 MHz

Channel 2: 2 mVpp – 3 Vpp into 50 Ω

4 mVpp – 6 Vpp into open circuit

Amplitude Resolution

Up to 4 digits

Amplitude Accuracy

(100 kHz)

± (0.3 dB + 1 mVpp of setting value)

Amplitude Flatness

± 0.3 dB

Advertising