Installation – Hach-Lange GANIMEDE N User Manual

Page 18

Advertising
background image

18

Installation

Figure 10

Use of the carrying strap

1. Place the analysis unit on a stable and level tabletop. Level out any unevenness with

the rear adjustable feet.

2. Remove the transport locking screws on the bottom of the sample changer.

To do so, tip the sample changer to the side and remove the screws above the
corresponding sign.

3. Position the sample changer to the left of the analysis unit on a stable and level

tabletop.

4. Place the open-loop control unit near (approx. 0.5 m 1.64 ft) to the analysis unit.

Note: Communication between the analysis unit and the open-loop control unit takes place via
IR interfaces.

1

Slip the carrying strap over the analysis unit as shown in
the illustration.

3

Tightly pull the carrying belt upwards at the front and
back sides.

2

Carefully tip the analysis unit to the side and guide the
outer parts of the carrying strap behind the feet of the
unit. Proceed in the same way on the opposite side.

4

The analysis unit must now be transported safely over a
short distance by two people.