Ganimede P analysis unit

Type

Ganimede P for the automatic serial analysis of orthophosphate
and total phosphate in the analytical laboratory, consisting of
analysis unit, sample changer and control unit

Photometric method

By means of ammonium molybdate in accordance with DIN EN
1189 of December 1996

Total P digestion

Rapid digestion at 150 °C and 6 bar

Sample pH

pH 2 – 11

Measurement range

0.01 – 3.8 mg/L PO

4

-P

Measurement accuracy

±0.03 mg/L at 1 mg/L

Measurement wavelength

880 nm

Calibration

2-point calibration and 10-point calibration

Processing time per sample

4 minutes for single determination

Ganimede N analysis unit

Type

Ganimede N for the automatic serial analysis of inorganically and
organically bound nitrogen in the analytical laboratory, consisting
of analysis unit, sample changer and open-loop control unit

Photometric method

By means of UV self-absorption at 210 nm and 228 nm in
accordance with DIN EN ISO 11 905-1

Digestion

Alkaline digestion at 150 °C and 8 bar

Measurement range

Low measurement range 0.5 – 30 mg/L TN

B

High measurement range 30 – 150 mg/L TN

B

Reproducibility of the measurement results

< 2 % at 15 mg/L

Carryover

< 1 % within a measurement range

Measurement wavelengths

210 nm and 228 nm

Calibration

2-point calibration and multiple-point calibration
(2–11 measurement points)

Processing time per sample

7 minutes for single determination

Sample changer

Sample tray

Sample tray for 36 × 25 mL or
Sample tray for 53 × 15 mL
Tray diameter 240 mm

Environmental conditions

Temperature range

15 – 35 °C

Relative humidity

< 75 %, non-condensing

Air pressure

75 – 106 kPa

Equipment properties

Particle mobility

Analysis unit

According to DIN 38402; the prerequisites of EN 1484 TOC of
> 100 µm are significantly exceeded

Sample changer

According to DIN 38402; the prerequisites of EN 1484 TOC of
> 100 µm are significantly exceeded with 1 mm (sample needle)