Installation

For a two-point calibration, the standards are positioned with the measurement series
on a sample tray for 36 sample containers as follows:

Position 33: 1st low standard, e.g. 0.5 mg/L TN

B

GCZ 100

Position 34: 2nd low standard, e.g. 30 mg/L TN

B

GCS 050

Position 35: 1st high standard, e.g. 30 mg/L TN

B

GCS 050

Position 36: 2nd high standard, e.g. 100 mg/L TN

B

GCS 100

The low standards can be selected freely in the range 0.5–30 mg/L TN

B

; the high

standards can be selected freely in the range 30–150 mg/L TN

B

.

Single and double determinations can be carried out. For double determinations, the
sample is drawn from a container twice. The settings refer to the whole sample tray.

The following table and illustration show an example of a sample tray load for a two-point
calibration:

Table 5 Sample tray load for 36 sample containers, Ganimede N

Position

Load

33

1st low standard, e.g. 0.5 mg/L TN

B

GCZ 100

34

2nd low standard, e.g. 30 mg/L TN

B

GCS 050

35

1st high standard, e.g. 30 mg/L TN

B

GCS 050

36

2nd high standard, e.g. 100 mg/L TN

B

GCS 100

1

Sample 1

2

Sample 2

3

Sample 3

4

Sample 4

5

Sample 5

6 etc.

Sample 6 etc.