Appendix a interface protocol – Hach-Lange GANIMEDE N User Manual

Page 59

Advertising
background image

59

Appendix A Interface protocol

RS 232 interface for the analysis unit and open-loop control unit.

The RS 232 C interface has a 9-pole sub-D socket and a 9-pole sub-D connector. The
interface is connected to the serial interface (COM) of the computer (LIMS system) with
the PC cable (XLH 917).

Table 6 RS 232 I/O card connection assignment

9-pole SUB-D

Socket

9-pole SUB-D

Connector

PIN

PIN

GND

5

GND

5

RxD

3

RxD

2

TxD

2

TxD

3

RTS

8

RTS

7

CTS

7

CTS

8

Table 7 RS 232 software description

ID Description

ID001

Start of a block

ID101

Parameters

ID102

Name

ID103

Users

ID104

Measurements (single/double)

ID107

Calibration (two-point/e.g. standard curve 1)

ID108

Sample tray measurement data

ID110

Method 0 name (total/low)

ID111

Standard absorbance 2 (two-point calibration)

ID112

Standard concentration 2 (two-point calibration)

ID114

Standard curve linear equation

ID115

Standard curve measurement data

ID116

Standard absorbance 1 (two-point calibration)

ID117

Standard concentration 2 (two-point calibration)

ID118

Standard curve lowest concentration

ID119

Standard curve highest concentration

ID120

Method name 1 (ortho/high)

ID121

Standard absorbance 2 (two-point calibration)

ID122

Standard concentration 2 (two-point calibration)

ID124

Standard curve linear equation

ID125

Standard curve measurement data

ID126

Standard absorbance 1 (two-point calibration)

ID127

Standard concentration 2 (two-point calibration)

ID128

Standard curve lowest concentration

ID129

Standard curve highest concentration

ID130

Start of a sample

ID131

Position

ID132

Sample name

Advertising