43

Operation

4. Check the capacity of the waste bottle.

5. Start the rinse procedure with the RINSE shortcut key.

6. Switch on the prepared analysis unit.

7. Confirm the safety prompt.

A connection to the analysis unit is established automatically via the IR interface.

8. Select the desired parameter.

The rinse process is started. The rinsing of the analysis unit is finished after two
minutes and the system can be switched off.

4.2.5

Stop

In the event of damage, an error or an incorrect load, the analysis series can be
interrupted immediately with the STOP shortcut key. A connection to the analysis unit is
established automatically via the IR interface.

1. Select and confirm the desired parameter.

The measurement series is interrupted.

2. Close the window with the upper menu key.

If the error or damage cannot be fixed without switching off the analysis unit, proceed as
follows:

3. Replace the reagent bottles with the cleaning solutions and activate RINSE.

4. Connect the reagent bottles with the analysis unit again for further analyses.

5. Load the sample tray.

6. Define the sample batch.

7. Start with ANALYZE.

4.2.6

Data/current

After the connection has been established, a list of the latest current measurement results
is shown with the Data/CURRENT menu option. In addition to the mean value, both single
results and single absorbances are displayed for a double determination. Measurement
results that are beyond the selected measurement range are shown in red.

1. Select DATA and CURRENT.

A connection to the analysis unit is established.

2. Select the desired parameter.

The list of current results opens.

3. Select the result to be edited with the cursor keys and confirm with the enter key.

The measurement results can now be edited.

The sample name, the unit, the indication of the result, e.g. as PO

4

, PO

4

-P or P

2

O

5

,

the sample collection time and the note can be changed.

If the dilution factor or the sample-specific factor is changed, the result is calculated
again automatically.

The position, result, method and measurement time components cannot be changed
afterwards.

4. Select CALIBRATION DATA in order to display the calibration that was carried out.