Installation

Table 1 Tube overview for analysis unit P

3.1.4.2

Installation of the tubes for the Ganimede N

1. Install all tubes of the analysis unit in accordance with

Figure 12 Tube plan for

analysis unit N, page 25

.

2. Screw the large blue lid onto the empty waste bottle (1500 mL).

Note: The waste bottle can take the reagent and sample waste from 100 individual
determinations.

3. Connect the GCA 200 reagents with the corresponding lids of the same color.

4. Install sample tube A.

To do so, lead the free end of sample tube A out through the left-hand side of the
analysis unit housing.

Connect the upper end of the sample needle with sample tube A.

5. Insert the sample needle into the guide of the sample changer arm.

Note: The sample needle must not reach the bottom of the sample container when magnetic
rods are used.

Identification

Color in tube

plan

Length

Dia-

meter

Colored

adapter

Connected:

A: (included in

the scope of

delivery of the

sample

changer)

Black

400 mm

2.67 mm

-

Sample changer sample needle
-> Left-hand side of the analysis unit housing
-> Valve 1 connection f

B

Red

140 mm

2.67 mm

-

Valve 1 connection b
-> Left-hand analysis cuvette

C

Pink

380 mm

3.04 mm

Yellow/blue

Valve 1 connection g
-> Pump 1
-> Digestion cuvette

D

Gray

300 mm

3.04 mm

-

Digestion solution A
-> Valve 1 connection d

E

Green

530 mm

3.2 mm

Yellow/blue

Right-hand analysis cuvette
-> Pump 2
-> Valve 2 connection g

F

Blue

450 mm

3.2 mm

-

Color solution B
-> Valve 2 connection e

G

Green

300 mm

3.04 mm

-

Reduction solution C
-> Valve 2 connection c

H

Orange

180 mm

3.04 mm

-

Distilled water
-> Valve 2 connection b

J

Black

550 mm

3.53 mm

Gray/green/g

reen

Analysis cuvette
-> Pump 3

K

Purple

500 mm

4.17 mm

Gray/black/p

urple

Lower analysis cuvette
-> Pump 4
-> T-piece -\> Valve 1 connection c
and -\> Waste bottle

L

Black

300 mm

3.04 mm

-

Acid solution D
-> Valve 2 connection d