Installation

3.1.3

Installation of the cables

1. Select the appropriate mains voltage (230 V or 115 V) for the region with the voltage

selector switch on the rear side of the analysis unit.

2. Connect the analysis unit to an earthed mains socket with the power cable (230 volts

+5 %/-15 %/50 Hz).

3. Optionally, connect a printer or PC to the corresponding interface of the analysis unit

4. Connect the SAMPLER interface of the analysis unit to the sample changer interface

with the connection cable (XLH917).

5. Connect the sample changer to the "sample changer power connector" on the

analysis unit with the power cable.

Note: Use only the supplied power cable.

6. Connect the external power supply unit of the open-loop control unit to an earthed

mains socket (230 volt +5 %/-15 %/50 Hz).

Note: Use only the external LZV563 power supply unit.

Note: The open-loop control unit is equipped for rechargeable battery operation. The charge
status is indicated permanently on the display. A red battery symbol indicates a weak
rechargeable battery. The battery symbol is shown in black when the rechargeable battery is
being charged. The fully charged rechargeable battery enables an operating time of
approximately 2.5 hours.

D A N G E R

Danger of an electric shock.
Use only the supplied power cable.
If another power cable without a protective earth contact is used or if the sample changer is
connected to a mains connection without a ground connection, there is the danger of an electric
shock for the operator and of the destruction of the device.

WA R N I N G

Electrical and fire hazards.
Use only the supplied power cable.
Only qualified experts may perform the tasks described in this section of the manual, while
adhering to all locally valid safety regulations.

WA R N I N G

Removable power cables must not be replaced with inadequately dimensioned power cables.

N O T I C E

Use only an earthed socket for the connection of this device to the power supply.
If you are not sure if the socket is earthed, have this checked by a qualified electrician.
The power plug serves in addition to the power supply to isolate the device quickly from the
mains where necessary.
This is recommended when the instrument is not used for a long time, and can prevent potential
dangers in the event of a fault.
Therefore, make sure that socket to which the device is connected is easy to reach by each user
at all times.