Maintenance – Hach-Lange GANIMEDE N User Manual

Page 48

Advertising
background image

48

Maintenance

8. Clean the cuvette with a moist cotton bud from inside.

9. If necessary, use a suitable cleaning solution, e.g. Extran for Ganimede P, alcohol for

Ganimede N (ethanol, propanol).

Note: Do not use any cleaning agents that could contain detergents on the Ganimede N.

10. Rinse the cuvette with distilled water or allow the cleaning agent to evaporate fully.

11. Ganimede P:

Insert the cleaned cuvette and pull the cuvette downward completely.

Ganimede N:

Insert the cleaned cuvette so that the guide rail on the left-hand side of the cuvette is
flush with the upper edge of the optical window.

12. Attach the cuvette carefully with the plastic screws.

13. Attach the connection pieces with the screws.

14. Connect all tubes according to

Figure 18

or

Figure 19

.

Figure 18

Cleaning of the Ganimede P analysis cuvette

Tube B

Tube J

Tube E

Tube K