4 [+signals] menu, 1 overview of +signals menu, 5 [+op.counters] menu – Hach-Lange FILTRAX eco User Manual

Page 33: 1 overview of +op.counters menu, 6 [+service] menu, Operation

Advertising
background image

33

Operation

4.4 [+SIGNALS] menu

The [+SIGNALS] menu is used for services and manufacturer
inspections. It is not required for operation or the adjustment of
instrument settings.

Note: Normal operation continues while the instrument is in this menu. If
no buttons are pressed for over ten minutes, the instrument automatically
switches to the operation display.

4.5 [+OP.COUNTERS] menu

The [+OP.COUNTERS] menu provides information on the
operating time of the various instrument components and wear
parts. It is not required for operation or the adjustment of instrument
settings.

Note: Normal operation continues while the instrument is in this menu. If
no buttons are pressed for over ten minutes, the instrument automatically
switches to the operation display.

4.6 [+SERVICE] menu

Important Note: With the exception of purely visual inspection, all
maintenance and service work must be carried out via this menu.

Note: When the menu is called up, the message "OK -> NO SAMPLE!" is
shown first of allto signify that the activation of this menu interrupts the
supply of samples to the measuring instruments immediately. Operation
only recommences following a definitive exit from the menu via the
keyboard or a called up program.

4.4.1 Overview of +SIGNALS menu

+SIGNALS

Setting

Description

TEMPERATURE (housing)

Information only

Temperature in the control unit in °C or °F

4.5.1 Overview of +OP.COUNTERS menu

+OP.COUNTERS

Description

CLEANED in the

Month when last clean took place, automatically updated via [FILTERCLEANING]

MODUL

Operating time of the filter module in days

P.TUBES IN

Remaining operating time for pump tube set in days (91 - 0 - negative number

1

)

P.CARTR. IN

Remaining operating time for pump cassette in days (365 - 0 - negative number

1

)

P.ROLL. IN

Remaining operating time for pump rollers in days (365 - 0 - negative number

1

)

AIR F.H. IN

Remaining operating time for the two air filters in the control unit in days
(91 - 0 - negative number

1

) — the air filter on the air outlet side may not necessarily need to

be replaced that often

1

When replacement date is exceeded.