43

Maintenance

5.3 Instrument decommissioning and storage

For outdoor installations, the control unit should remain
permanently in operation to make sure the climate control is
protected against heat and frost.

5.3.1 Filter module

If the system is decommissioned:

1. Remove the filter module from the basin.

2. Remove the sample tube.

3. Remove the filter module from the module carrier.

4. Clean the filter module

(refer to

Section 5 Maintenance on page 35

).

5. Pack the filter module in a plastic bag (EYV017) to protect it

from drying out.

5.3.2 Control unit

6. Select the [+SERVICE] menu and confirm.

7. Remove the suction tube and sample tube (refer to section

3.3.3 Install suction tube on page 20

).

8. Select the [PREPUMP] menu and confirm to allow the internal

tube system to drain.

9. Select the [+SERVICE] menu and confirm.

10. Disengage the pump cassette (refer to

Figure 23 Pump tube

replacement on page 38

).

11. Disconnect the system from the mains.

12. Uninstall the control unit and store in a dry location.

SERVICE-Menu

Ok NO SAMPLE!

Ok

PREPUMP

Ok

PREPUMP 60 s

2x