Section 1 specifications – Hach-Lange FILTRAX eco User Manual

Page 5

Advertising
background image

5

Section 1

Specifications

Subject to change without notice

Electrics

Power supply

230 V (Option: 115 V) ±10 % AC voltage, 50/60 Hz

Power consumption

Device with system components: 1500 VA

Line

Heater off

Max.

(–20 °C [–4 °F])

start-up

Cont.

(–20 °C [–4 °F])

2 m [6.56 ft]
10 m [32.8 ft]
20 m [65.6 ft]
30 m [98.4 ft]

150 VA
150 VA
150 VA
150 VA

450 VA
950 VA

1500 VA
2100 VA

< 200 VA
< 300 VA
< 400 VA
< 500 VA

Outputs
(Use screened cables!)

Fault contact:

potential-free contact (230 V, max. 3 A)

Alarm contact:

potential-free contact (230 V, max. 3 A)

Service interface: RS 232

Protection type (control unit)

IP55

Fuses

T 4A E; 250V (2 x)
T 7A E; 250V (2 x)

Environment

Medium temperature

+5 °C to +40 °C [41 °F to 104 °F]

Ambient temperature

–20 °C to +40 °C [–4 °F to 104 °F]

Sample quantity

Approx. 600 mL/h for up to 2 process photometers

Delivery height

Module carrier — control unit: 3 m [9.84 ft]
Control unit — process measuring instrument: 7 m [22.97 ft]

General specifications

Line length

Suction tube (heated):

5 m [16.4 ft]

Sample tube (unheated):

2 m [6.56 ft]

Sample tube (heated):

2 m [6.56 ft], 10 m [32.8 ft], 20 m [65.6 ft],
30 m [98.4 ft]

Maintenance requirements

Approx. 1 hour/month

Weight

Control unit:

approx. 20 kg [44 lb]

Module carrier with 5 m [16.4 ft] suction tube: approx. 3.5 kg [77 lb]
2 m [6.56 ft] sample tube:

approx. 1 kg [2.2 lb]

10 m [32.8 ft] sample tube:

approx. 5 kg [11 lb]

20 m [65.6 ft] sample tube:

approx. 10 kg [22 lb]

30 m [98.4 ft] sample tube:

approx. 15 kg [33 lb]

Dimensions (W × H × D)

Control unit:

430 Ч 530 Ч 220 mm [16.9 Ч 20.86 Ч 8.66 in.]

Filter module:

251 Ч 445 Ч 17 mm [9.9 Ч 17.5 Ч 0.67 in.]

Certification

Certification

CE. The manufacturer clarifies conformity with the applicable EU safety
guidelines and EMC guidelines.

Advertising