Operation

When this menu is selected,

the pump stops immediately

the red indicator light comes on

the fault relay is set

Note: The climate control and heater systems for the control unit and the
tubes remain active.

Note: The functions of the individual menu items and submenus
([FILTERCLEANING], [OP.COUNTERS], [+TEST OF FUNCT.]) are
explained in detail in

Section 5 Maintenance on page 35

and section

6.1 Messages on page 45

.

They are not required for normal operation.

4.6.1 Overview of +SERVICE menu

+SERVICE

Setting

Description

EXTRACTION

50 to 130%
(Initial setting 100% = 600 mL/h, this
sample size is produced under normal
installation and operating conditions
with clean filter modules)

The delivery volume of the pump is set by the
speed.

PREPUMP

The pump is activated for ten minutes in order to fill
all the tubes. Operation is then resumed
automatically.

FILTERCLEANING

Menu-controlled cleaning is described in detail in
section

5.2.1 Menu-controlled cleaning tasks on

page 39

. After each clean, the date in the

operating time counter [CLEANED] is
automatically updated and operation is resumed.

+OP.COUNTERS

Calls up the [+OP.COUNTERS] menu — all
counters can now be reset.

+TEST OF FUNCT.

Component(s) Options
Pump

(3 s forward and
back), stop

Fan housing

on, off

Heater housing on, off
Heater line

on, off

Warning

on, off

Malfunction

on, off

Indicator lights on, off

(green/red at same
time)

Each of these components can be controlled and
tested individually.

+SIGNALS

Calls up the [+SIGNALS] menu.

+DEVICEDATA

Calls up the [+DEVICEDATA] menu.

Status

Detailed description of the error and
acknowledgement

Advertising