39

Maintenance

5.2 Cleaning

Normal operation of this device may require the use of samples that
are biologically unsafe.

5.2.1 Menu-controlled cleaning tasks

D A N G E R

Potential danger with contact with chemical/biological substances.
Working with chemical samples, standards and reagents can be
dangerous.
Make yourself familiar with the necessary safety procedures and the
correct handling of the chemicals before use and read and follow all
relevant safety data sheets.

WA R N I N G

Observe all cautionary information printed on the original solution
containers and safety data sheets prior to their use.
Dispose of all consumed solutions in accordance with the local and
national regulations and laws.
Select the type of protective equipment suitable to the concentration and
quantity of the dangerous material being used.
Observe instructions and safety information (information on risks and
safety) on the containers for the chemicals used.
Wear safety clothing: Lab coat, Safety glasses, Rubber gloves

Table 2 Menu-controlled cleaning tasks

Action

FILTRAX eco display

1

For all connected process measuring instruments and for the FILTRAX eco, call up the
[+SERVICE] menu.
Next, call up the [+FILTERCLEANING] menu.

+FILTERCLEANING
NO SAMPLE!

2

Lift the module carrier out of the basin or channel and confirm with [OK].

REM.MOD.CARRIER

3

Detach the sample tube from the module carrier, disconnect the filter module from the
module carrier and confirm with [OK].

REMOVE MODUL

N O T I C E

Do not touch the sensitive membrane and never allow it to dry out. Only store the filter module
in the plastic bags provided.