Troubleshooting

6.3 Malfunctions

Text is displayed — the red indicator light is lit and the fault relay
contact is set.

6.4 Voltage drop (power failure)

If the power supply is interrupted (e.g. due to a power failure), no
indicator light is lit. The warning relay contact and fault relay contact
are automatically closed.

N O T I C E

The instrument may be damaged. Take appropriate counteractive
measures as soon as possible.

Table 7 Malfunctions

Malfunctions displayed

Definition

Resolution

WARM UP MODE!

The temperature in the housing is < 1 °C [34 °F],
the instrument is warming up

Wait until the heater has raised the
temperature inside the housing to > 1 °C
[34 °F] — this warm-up phase then takes
another 30 minutes. Keep the door closed
during this time.

DEVICE ERROR

Serious electronic error/bus error.
Normal operation no longer possible.

Unplug from the mains for 1 minute.
If the error is still present: Call customer
service.

TEMP.SENSOR

Temperature sensor faulty or not installed
Note: Normal operation is stopped because it is
not possible to regulate the temperature.

Check the temperature sensor connection.
If the error is still present: Call customer
service.