Nova-Tech pH-CON 10 User Manual

Page 10

Advertising
background image

Instruction Manual

PC 10

- 6 -

4 . C A L I B R AT I O N

4.1.

Important Information on Meter Calibration

When you recalibrate your meter, old pH and conductivity calibration data are replaced on a

point by point basis. For example, if you previously calibrated your meter at pH 4.0, 7.0, and

10.0, and you recalibrate at pH 7.0, the meter retains the old calibration data at pH 4.0 and pH

10.0.

To completely recalibrate your meter, or when you use a replacement probe, it is best to set

the meter to its factory defaults and recalibrate the meter at all points.

To reset the meter to its factory defaults:

1. While in Measurement mode, press

CAL/MEAS and hold for 3 seconds.

2.

The meter will prompt RST in the lower display

and CAL in the upper display. See Figure 6.

3.

Press ENTER to reset your meter to its factory

defaults. If you do not want to erase existing

calibration data, press CAL/MEAS to escape

this mode.

For information on how to calibrate your meter:

See Section 4.3 on page 8 for Temperature calibration.

See Section 4.4 on pages 10 for pH calibration.

See Section 4.5 on pages 12 for conductivity calibration.

Figure 6: Reset to factory defaults