Nova-Tech pH-CON 10 User Manual

Page 11

Advertising
background image

Instruction Manual

PC 10

- 7 -

4.2.

Preparing the Meter for Calibration

The pH/conductivity/temperature probe included with this meter is designed for use with this

meter only. Do not substitute other types of probes or electrodes. For a replacement probe,

see the “Accessories” section, page 21.

Be sure to remove the protective cap of the pH probe before calibration or

measurement. Always immerse the probe beyond the electrode band.

Immerse the probe in tap water for 10 minutes before calibrating or taking readings to hydrate

the pH electrode surface and minimise drift.

Rinse your probe in DI water after use. Storage in electrode storage solution is ideal.

Alternatively, pH 4.0 or 7.0 calibration standards can be used for short term storage. Do not

reuse buffer solutions after calibration. Contaminants in the solution can affect the calibration,

and eventually the accuracy of the measurements.

Always immerse the
electrode beyond this band

Always remove this protective cap
before calibration or measurement