Nova-Tech pH-CON 10 User Manual

Page 12

Advertising
background image

Instruction Manual

PC 10

- 8 -

4.3. Temperature

Calibration

The built-in temperature sensor included in the

probe is factory calibrated. Calibrate your sensor

only if you suspect temperature errors may have

occurred over a long period of time or if you have a

replacement probe.

1.

Turn the meter on. Press MODE to select pH

Measurement mode.

2. Press the CAL/MEAS key to enter pH

calibration mode. The CAL indicator will appear

above the primary display.

See Figure 10.

3.

While in pH calibration mode, press the MODE

key to enter temperature calibration mode. The

primary display shows the temperature reading

with zero offset and the secondary display

shows you what the temperature value was

initially.

See Figure 11.

4. Dip the probe into a solution with known

temperature (i.e. a temperature bath). Allow

sufficient time for the meter’s temperature

reading to stabilize.

5. Press

the

S or T key to set the primary display

reading to the correct temperature value. (i.e.

the temperature of the temperature bath)

See Figure 12.

Figure 7: Enter pH Calibration mode

Figure 8: Initial temperature value

Figure 9: Adjust temperature value