7 . P R O B E C A R E A N D M A I N T E N A N C E

Under typical operating conditions, the probe will need to be replaced every 6 to 12

months. In extreme applications, the probe may wear out sooner. Proper care and

maintenance will help you receive the maximum probe life and ensure more accurate

readings.

Keep the probe clean. Before use, rinse the probe twice. Be sure to remove the protective

cap of the pH probe before use and always immerse the probe beyond the electrode band.

For the best accuracy, soak your probe in electrode storage solution, pH 4.0 buffer, or tap

water for at least 5 to 10 minutes before calibration.

Gently stir the probe in the solution while you take readings. Tap the probe gently against the

bottom and sides of your container to remove any air bubbles, which may interfere with

accuracy.

Clean the probe thoroughly by immersing it in an agitated mild detergent bath. To clean the

conductivity cell, remove the plastic sleeve and use a cotton swab soaked in isopropyl alcohol

to clean the steel pins. Wipe the steel pins if needed with soft tissue paper. Replace the

sleeve which is necessary for calibration and measurement. DO NOT wipe the pH bulb as it

may cause static build-up. After cleaning the probe, rinse the probe well with clean water.

Recalibrate the meter after cleaning the probe.

Do not strike the probe against any hard surface. Do not immerse the probe in oily solutions.

Store the probe in its cap filled with electrode storage solution. To order a replacement probe

and other accessories, see “Accessories” section on page 21 .