Nova-Tech pH-CON 10 User Manual

Page 16

Advertising
background image

Instruction Manual

PC 10

- 12 -

4.5. Conductivity

Calibration

The Important Information on Conductivity Calibration

Your meter has four measuring ranges. You can calibrate one point in each of the measuring

ranges (up to four points). If you are measuring values in more than one range, make sure to

calibrate each of the ranges you are measuring.

The following table lists the corresponding conductivity range. You should calibrate each

range using a solution that falls between the values in the “recommended calibration solution

range” column.

Conductivity Range

Recommended Calibration

Solution Range

0.00 to 19.99

PS

6.00 to 17.00

PS

0.0 to 199.9

PS

60.0 to 170.0

PS

0 to 1999

PS

600 to 1700

PS

0.00 to 19.99 mS

6.00 to 17.00 mS

NOTE:

Minimum calibration value should be 20% of full scale for the specific range. E.g.: The

LOWEST calibration value for the range of 0.00 to 19.99

PS is 4.00 PS.

When you recalibrate your meter, old calibrations are replaced on a range basis. For example,

if you previously calibrated your conductivity meter at 1413

PS in the 0 to 1999 PS range and

you recalibrate at 1500

PS (also in the 0 to 1999 PS range), the meter will replace the old

calibration data (1413

PS) in that range. The meter will retain all calibration data in other

ranges.

To completely recalibrate your meter, or when you use a replacement probe, it is best to clear

all calibration data in memory. To erase all the old conductivity data, see Section 4.1:

Important Information on Meter Calibration.

NOTE:

These meters are factory set to a temperature coefficient of 2.00% per °C. For most

applications this will provide good results. The factory default value for normalization

temperature is 25°C.