Nova-Tech pH-CON 10 User Manual

Page 13

Advertising
background image

Instruction Manual

PC 10

- 9 -

6. Press the ENTER key to confirm temperature

calibration. The CON indicator will flash to

confirm calibration.

See Figure 13.

NOTES

To exit from Temperature Calibration mode without

confirming calibration, DO NOT press ENTER in step

6. Press CAL/MEAS instead.

Temperature calibration is restricted to ±5°C from

the initial value displayed during calibration (shown

in the secondary display).

Since temperature readings affect the accuracy of the pH and conductivity measurements, it is

strongly recommended to carry out pH and conductivity calibration after a temperature

calibration is done.

Figure 10: CONFIRM temperature

calibration

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682