Nova-Tech pH-CON 10 User Manual

Page 19

Advertising
background image

Instruction Manual

PC 10

- 15 -

5 . M E A S U R E M E N T

The READY indicator appears on the display when

the readings stabilise. It will turn off if the readings

start to fluctuate.

NOTE: Be sure to remove the protective cap of the

pH electrode before measurement.

To take readings:

1. Rinse the probe with DI water before use to

remove any impurities adhering to the probe

body. If the pH electrode has dehydrated,

soak it for 30 minutes in tap water.

2.

Switch on the meter. The MEAS annunciator

appears on the top center of the LCD. The

ATC indicator appears in the lower right hand

corner to indicate Automatic Temperature

Compensation. See Figure 22.

3.

Dip the probe into the sample.

NOTE: When measuring a sample, make sure that

the probe is immersed beyond the electrode band.

See figure in Section 4.2. Stir the sample using the

probe gently to create a homogenous sample. Allow

time for the readings to stabilize. Note the readings

on the display. When the readings are stable, the

READY annunciator appears.

4. To toggle between pH and conductivity

readings, press the MODE key. See Figures

23 and 24.

NOTE: Conductivity readings are auto

ranging and will automatically move to

the correct range (four ranges

possible).

MEAS

pH

°C
ATC

Figure 19: Measurement mode

Figure 20: READY

Figure 21: Conductivity mode