Nova-Tech pH-CON 10 User Manual

Page 15

Advertising
background image

Instruction Manual

PC 10

- 11 -

6.

After the reading has stabilized and the READY

indicator is displayed, press ENTER to confirm

calibration. A confirming indicator (CON)

flashes and disappears. The meter is now

calibrated to the buffer indicated in the

secondary display.

See Figure 13.

The secondary display automatically scrolls to the

next buffer calibration option.

If you are performing multi-point calibration, go to

step 7.

If you are performing one-point calibration, go to step

10.

7. Press

the

S or T key to select the next buffer value you want to calibrate (pH 4.00, 7.00

or 10.00).

See Figure 14.

8. Rinse the probe with DI water or a rinse

solution, and place it in the next pH buffer you

want to calibrate.

9.

Follow steps 5 to 8 for all calibration processes

(up to 3 points).

10.

When calibration is complete, press

CAL/MEAS to return to pH measurement

mode.

NOTES

To exit from pH Calibration mode without confirming calibration, DO NOT press ENTER in

step 6. Press CAL/MEAS instead.

If the selected buffer value is not within ±1.00 pH from the measured pH value:

The electrode and buffer icon blink and the ERR annunciator appears in the lower left corner

of the display. These indicators also flash if the buffer used is not the same as the buffer value

on the secondary display.

Figure 13: CONFIRM calibration point

Figure 14: Next calibration point

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682