4.6.

Calibrating for Conductivity:

1.

Pour out two separate portions of your calibration

standard and one of DI water into separate clean

containers.

2. If necessary, press MODE key to select

Conductivity Mode. The

PS and mS indicator will

appear on the right side of the display.

3. Rinse your probe with DI water, then rinse the

probe in one portion of the calibration standard.

Immerse the probe into the second portion of the

calibration standard. The meter auto ranging

function selects the appropriate conductivity range

(four ranges are possible). Be sure the probe is

free from air bubbles and sufficient calibration

standard is used to cover the conductivity sensor.

Wait for the reading to stabilize. The READY

indicator lights when the reading is stable.

See Figure 18

4. Press the CAL/MEAS key. The CAL indicator

appears above the primary display. The primary

display shows the measured reading and the

secondary display shows the temperature.

See Figure 19.

5.

Press the ź or Ÿ key to scroll to the value of your

conductivity standard. Press and hold the ź or Ÿ

keys to scroll faster. See Figure 20.

MEAS

READY

°C
ATC

mS

CAL

READY

°C
ATC

mS

CAL

READY

°C
ATC

mS

Figure 16: CAL Indicator

Figure 15: Stabilized measurement

Figure 17: Adjust to standard value