Teledyne 6400A - Sulfur dioxide analyzer User Manual

Page 106

Advertising
background image

TAI Model 6400A SO

2

Analyzer Operator Manual, 02164, Rev. G

7-24

Table 7-15: NIST-SRM's Available for Traceability of Calibration and Audit Gas

Standards

NIST-SRM

4

Type

Nominal concentration

1693

1694

SO

2

in N

2

SO

2

in N

2

50 ppm

100 ppm

Permeation Tubes

NIST-SRM

4

Type

Permeation rate,
ug/min @ 25

o

C

Concentration, ppm at flow rates
of: 1 L/min / 5 L/min

1625

1626

1627

SO

2

permeation tube

SO

2

permeation tube

SO

2

permeation tube

2.8

1.4

0.56

1.07 / 0.214

0.535 / 0.107

0.214 / 0.0428

Cylinders of working gas traceable to NIST-SRM's (called EPA Protocol Calibration Gas) are
also commercially available (from sources such as Scott Specialty Gases, etc.).

7.9 Special Calibration Requirements for Dual Range or

Auto Range

If Dual Range or Auto Range is selected, then it should be calibrated for both Range1 and Range2
separately. Pressing CAL key will prompt #1 and #2 keys for Range1 (Low Range) or Range2 (Hi
Range) calibration selection. Select desired range number and press ENTR to proceed to the
calibration. Once Range# is selected, the display will show Test Measurements and SO

2

concentration for the corresponding Range#. You must enter expected SO

2

gas concentrations

separately per Table 7-3 procedure for each Range #.

For zero calibration allow zero gas through the sample port and proceed to manual zero calibration
procedure per Table 7-2, step 2 through step 4. After zero calibration is set, switch to span SO

2

gas

to continue for span calibration procedure per Table 7-4 step 2 through step 4. Press EXIT to exit
from the current Range.

Repeat above procedure for other Range # by pressing CAL key and selecting Range # as described
above. Enter once again corresponding SO

2

gas concentration for selected Range # and continue

zero/span calibration for other Range # selected.