TAI Model 6400A SO

2

Analyzer Operator Manual, 02164, Rev. G

7-15

Figure 7-2: Diagram Of Calibration System

Table 7-11: EPA Expected Span Gas Concentration Procedure

Step Number

Action

Comment

1. Press

CAL-

CONC

This key sequence causes the 6400A to prompt for the
expected SO

2

concentration.

Enter the SO

2

span concentration value by pressing the key

under each digit until the expected value is set.

2.

Press ENTR

ENTR stores the expected SO

2

span value.

3.

Press EXIT

Returns instrument to SAMPLE mode.

Sample the generated concentration until the SO

2

response is stabilized.

Span the instrument by the following procedure: