3.2 EPA Equivalency Designation
Teledyne Analytical Instruments, Model 6400A Sulfur Dioxide Analyzer is designated as Reference
Method Number EQSA-0495-100 as defined in 40 CFR, Part 53, when operated under the following
conditions:
Range: Any range from 50 ppb to 1000 ppb.
1. Ambient temperature range of 5 to 40
o
C.
2. Line voltage range of 105-125 VAC, 60Hz.
3. With 5-micron TFE filter element installed in the internal filter assembly.
4. Sample flow of 650 +/- 65 cc/min.
5. Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm or better.
6. Software settings:
A.
Dynamic
span
OFF
B.
Dynamic
zero
ON
or
OFF
C.
AutoCal ON
or
OFF
D.
Dual
range
ON
or
OFF
E.
Autorange
ON
or
OFF
F. Temp/Pressure compensation
ON
Under the designation, the Analyzer may be operated with or without the following options:
1. Rack mount with chassis slides (P/N 01469).
2. Rack mount without slides, ears only (P/N 01470).
3. Fluorocarbon zero/span valves (P/N 01491).
4. Internal zero/span (P/N 01441).
5. SO
2
Permeation tube - uncertified 0.4ppm @ 0.7 lpm (P/N 0150603)
6. SO
2
Permeation tube - certified 0.4ppm @ 0.7 lpm (P/N 0150604)
7. SO
2
Permeation tube - certified 0.8ppm @ 0.7 lpm (P/N 0150607)
8. SO
2
Permeation tube - uncertified 0.8ppm @ 0.7 lpm (P/N 0150608)
9. Zero air scrubber (P/N 01440).
4-20mA, isolated outputs (P/N 01471-1).