Teledyne 6400A - Sulfur dioxide analyzer User Manual
Page 14

TAI Model 6400A SO
2
Analyzer Operator Manual, 02164, Rev. G
2-2
2.
If desired, attach the analog output connections to a strip chart recorder and/or
datalogger. Refer to Figure 9-3 for jumper settings for the desired analog output voltage
range. Factory default setting is 0-5VDC.
3.
Connect the power cord to the correct voltage line, then turn to Section 2.3 Initial
Operation.
CAUTION
High voltages present inside case.
DO NOT LOOK AT THE UV LAMP, UV LIGHT
COULD CASE EYE DAMAGE.
ALWAYS USE SAFETY GLASSES
(PLASTIC GLASSES WILL NOT DO).
Connect the exhaust fitting on the rear panel to a
suitable vent outside of the room.
CAUTION
DO NOT OPERATE WITH COVER OFF.
Before operation check for correct line voltage and
frequency on Serial Number Sticker.
Do not plug in the power cord if the voltage or
frequency is incorrect.
Do not operate without proper chassis grounding.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or
connecting eletrical subassemblies.
Always replace shipping screws when transporting the Analyzer.