Teledyne 6400A - Sulfur dioxide analyzer User Manual
Page 25

TAI Model 6400A SO
2
Analyzer Operator Manual, 02164, Rev. G
2-13
Table 2-1: Final Test and Calibration Values (Continued)
Span and Cal Values
Parameter
Observed
Value
Units
Nominal Range
Reference Section
SO
2
Span Conc
PPB
20 - 20,000
Table 7-3
SO
2
Slope
1.0 ± 0.3
7.10
SO
2
Offset
mV
<250
7.10
Noise at Zero (rms)
PPB
0.1 - 0.2
Table 9-1
Noise at Span (rms)
PPB
0.5% of reading
(above 50 ppb)
Table 9-1
Measured Flows
Parameter
Observed
Value
Units
Nominal Range
Reference Section
Sample Flow
cc/min
650 ± 10%
9.2.2, 9.3.6,
Figure 9-6
IZS Purge Flow
cc/min
50 ± 10
6.3
Factory Installed Options
Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Fluorocarbon Zero/Span Valves
Internal Zero/Span - IZS
Permeation Tube (Output Specification)
External Pump
4-20 mA Current Loop Output
REC
DAS
SO
2
(RANGE 1)
SO
2
(RANGE 2)
SPARE
TEST OUTPUT
PROM #
Serial #
Date
Technician