Teledyne 6400A - Sulfur dioxide analyzer User Manual

Page 99

Advertising
background image

TAI Model 6400A SO

2

Analyzer Operator Manual, 02164, Rev. G

7-17

1. Zero calibration procedure is identical to Table 7-11.

2. Generate the SO

2

concentration to approximately 80% of the URL of the range.

Calculate the expected concentration as follows.

[

]

.

SO

P

F

F

c

2

3

4

4

10

382 10

=

Ч

+

Ч

Ч

--------------------------------------- Equation 7.2

[SO

2

] =

SO

2

gas concentration, ppm

P

= permeation flow rate at the specific temperature, ug·SO

2

/min.

F

c

= carrier flow rate over the permeation tube, SLPM (25

o

C, 760 mmHg)

F

d

= diluent air flow rate, SLPM (25

o

C, 760 mmHg)

Figure 7-3: Diagram F Permeation Calibration System