Analyzer Operator Manual, 02164, Rev. G

7-5

Enter the expected SO

2

span gas concentration:

Table 7-3: Enter Expected Span Gas Concentrations Procedure

Step Number

Action

Comment

1. Press

CAL-

CONC

This key sequence causes the 6400A to prompt for the
expected SO

2

concentration. Enter the SO

2

span concentration

value by pressing the key under each digit until the expected
value is set. This menu can also be entered from CALS.

2.

Press ENTR

ENTR stores the expected SO

2

span value.

3.

Press EXIT

Returns instrument to SAMPLE mode.

Table 7-4: Manual Span Calibration Procedure - Span Gas Through Sample Port

Step Number

Action

Comment

1.

Press CAL

The 6400A enters the calibrate mode. External span gas should
be fed to the sample port.

2.

Wait 10 min

Wait for reading to stabilize at span value

3.

Press SPAN

If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.

4.

Press ENTR

Pressing ENTR actually changes the calibration equations and
causes the instrument to read the SO

2

span concentrations.

5.

Press EXIT

The 6400A returns to sampling. Immediately after calibration,
data is not added to the DAS averages.