Table 2-1: Final Test and Calibration Values
TEST Values
Observed
Value
Units
Nominal Range
Reference Section
RANGE
PPB
50 - 20,000
5.3.4
STABIL
PPB
0.1 - 2
5.2.1, 9.1.1,
Table 9-1, 9.2.5
PRESS
in-Hg
25 - 35
9.1.1, 9.3.6,
Table 9-1
SAMP FL
CC / MIN
650 ± 10%
9.2.2, 9.3.6,
Table 9-1
PMT
mV
0 - 5000
9.4.1
UV LAMP
mV
2000 - 4000 typical
9.4.2
STR. LGT
PPB
<100
Table 9-1
DRK PMT
mV
-50 - +200
9.4.3, Table 9-1
DRK LMP
mV
-50 - +200
Table 9-1
SLOPE
1.0 ± 0.3
7.10
OFFSET mV
<250
7.10
HVPS
V
550 - 900 constant
9.3.10
DCPS
mV
2500 ± 200
9.3.5
RCELL TEMP
o
C
50 ± 1
9.3.7
BOX TEMP
o
C 8-50
9.3.4.1
PMT TEMP
o
C
7 ± 1
9.3.9
IZS TEMP
o
C
50 ± 0.3
9.5.4
Electric Test & Optic Test
Electric Test
PMT Volts
mV
2000 ± 1000
9.1.3.2
SO
2
Conc
PPB
1000 ± 500
9.1.3.2
Optic
Test
PMT Volts
mV
2000 ± 1000
9.1.3.3
SO
2
Conc
PPB
1000 ± 500
9.1.3.3
(table continued)