Section 4: Wiring

11/12

34

51182 Rev. 2

DATA RECORDING

1. Record scale serial number from the tag.

2. Record instrument, junction box and load cell serial numbers.

3. Keep a copy of the sheet in the customer file.

MOISTURE PROTECTION

Full Electronic Scales have been designed to provide protection from the effects of
moisture.

• The load cells are calibrated with the cable attached.
• The cable

MUST NOT

be cut.

• The cable is connected directly to the Junction Box through a sealed bushing

which

MUST BE TIGHTENED WITH PLIERS

to keep moisture out of the box.

• On all boxes, the black plastic fittings have O-rings that can be forced out of

position if the bushing itself is not tight.

─

To prevent this, first tighten the inner nut securing the bushing in the hole.

─

Then insert the cable and carefully tighten gland with pliers until it is very snug.

─

DO NOT over-tighten where bushing 'turns'.

─

All box covers MUST BE SECURED with ALL screws tightened properly (

18

to 20 lbs/in

) for protection against moisture.