Teledyne LeCroy PXA125 User Manual

Page 152

Advertising
background image

User Manual PXA125

Programming Reference 4-34

Purpose
This command is a high-level command that programs the PXA125 to output triangular waveform along with
its associated parameters

<freq> sets the output frequency in units of hertz

<ampl> sets the output amplitude in units of volts

<offs> sets the output offset in units of volts

<phase> sets the output start phase in units of degrees

Parameter type

Numeric

Parameter range

<freq>

10e3 to 50e6

<ampl>

80e-3 to 80

<offs>

-3.6 to +3.6

<phase>

0 to 360

APPLy:TRIangle?

Response
The PXA125 will return <freq>, <ampl>,<offs>,<phase>. Default values are

1e6,5,0,0

APPLy:SQUare{<freq>,<ampl>,<offs>,<duty_cycle>}

Purpose
This command is a high-level command that programs the PXA125 to output square waveform along with its
associated parameters

<freq> sets the output frequency in units of hertz

<ampl> sets the output amplitude in units of volts

<offs> sets the output offset in units of volts

<duty_cycle> sets the squarewave duty cycle in units of %

Parameter type

Numeric

Parameter range

<freq>

10e3 to 50e6

<ampl>

80e-3 to 8

<offs>

-3.6 to +3.6

<duty_cycle> 0 to 360

APPLy:SQUare?

Response
The PXA125 will return <freq>,<ampl>,<offs>,<duty_cycle>. Default values are

1e6,5,0,50

APPLy:PULSe{<freq>,<ampl>,<offs>,<delay>,<rise>,<high>,<fall>}

Purpose
This command is a high-level command that programs the PXA125 to output pulse waveform along with its
associated parameters

Advertising