Programming Reference 4-66

ARM:SLOPe{POSitive|NEGative}

Purpose
This command will set the edge sensitivity of which the PXA125 will stop in arm mode. The stop and start
commands are applied at the rear-panel TRIG IN connector.

•

POSitive will set the positive edge

•

NEGative will set the negative edge

Parameter type

Discrete

ARM:SLOPe?

Response and default
The PXA125 will return POS, or NEG depending on the present setting. Default value is POS.

ARM:BREakpoint:POSition<position>

Purpose
This command will program the breakpoint position for the arm function.

•

<position> will set the breakpoint position

Parameter type

Numeric (integer only)

Parameter range

<position>

0 to “n” in units of waveform points, where “n” is the size of the active segment. The
breakpoint position can be programmed in increments of 4 points minimum.

ARM:BREakpoint:POSition?

Response and default
The PXA125 will return the present breakpoint value. Default value is 0.

INITiate:CONTinuous{OFF|ON|0|1}

Purpose
This command will set the output in continuous operation and interrupted operation. The trigger subsystem
commands will affect the PXA125 only after it will be set to interrupted operation.

•

ON or 1 will set the continuous mode

•

OFF or 0 will set the interrupted mode

Parameter type

Discrete

INITiate:CONTinuous?

Response to query version
The PXA125 will return 1 if the output is continuous, or 0 if the output is interrupted. Default value is 0.