User Manual PXA125

Programming Reference 4-37

APPLy:GAUSsian{<freq>,<ampl>,<offs>,<exp>}

Purpose
This command is a high-level command that programs the PXA125 to output gaussian waveform along with
its associated parameters

<freq> sets the output frequency in units of hertz

<ampl> sets the output amplitude in units of volts

<offs> sets the output offset in units of volts

<exp> sets the exponent value

Parameter type

Numeric

Parameter range

<freq>

10e3 to 50e6

<ampl>

80e-3 to 80

<offs>

-3.6 to +3.6

<exp>

10 to 200

APPLy:GAUSsian?

Response
The PXA125 will return <freq>, <ampl>,<offs>,<exp>. Default values are

1e6,5,0,20

APPLy:DC{<%_ampl>}

Purpose
This command is a high-level command that programs the PXA125 to output DC level along with its
associated parameter

<%_ampl> sets the output level as percentage of the amplitude setting

Parameter type

Numeric

Parameter range

<%_ampl> -100 to +100 in units of %

APPLy:DC?

Response to query version
The PXA125 will return <%_ampl>. Default values are

100

APPLy:USER{<seg# >,<sclk>,<ampl>,<offs>}

Purpose
This command is a high-level command that programs the PXA125 to output user-defined waveform along
with its associated parameters

<seg# > selects the active segment number

<sclk> sets the sample clock frequency in units of samples per second

<ampl> sets the output amplitude in units of volts