User Manual PXA125

Programming Reference 4-52

Parameter range

<pulse_width> 0 to 99.9 in units of percent

PULSe:WIDTh?

Response and default
The PXA125 will return the present width value. Default value is 10.

PULSe:TRANsition<rise>

Purpose
This command programs pulse transition from low to high of the standard pulse waveform. This command
has no affect on arbitrary waveforms.

<rise> sets the rise time parameter

Parameter type

Numeric

Parameter range

<rise>

0 to 99.9 in units of percent

PULSe:TRANsition?

Response and default
The PXA125 will return the present rise time value. Default value is 10.

PULSe:TRANsition:TRAiling<fall>

Purpose
This command programs pulse transition from high to low of the standard pulse waveform. This command
has no affect on arbitrary waveforms.

<fall> sets the fall time parameter

Parameter type

Numeric

Parameter range

<fall>

0 to 99.9 in units of percent

PULSe:TRANsition:TRAiling?

Response and default
The PXA125 will return the present fall time value. Default value is 10.

RAMP:DELay<delay>

Purpose
This command programs delay of the standard ramp waveform. This command has no affect on arbitrary
waveforms.

<delay> sets the delay parameter