Teledyne LeCroy PXA125 User Manual
Page 188

User Manual PXA125
Programming Reference 4-70
•
OFF or 0 will turn off STAR triggers
Parameter type
Discrete
TRIGger:STAR?
Response and default
The PXA125 will return 1 if star lines are active, or 0 if star lines are disabled. Default value is 0.
TRIGger:TIMer<interval>
Purpose
This command will set the period for the internal trigger generator.
•
<interval> will set the timer of the internal trigger generator
Parameter type
Numeric
Parameter range
<interval>
100e-3 to 5e6 in units of seconds
TRIGger:TIMer?
Response and default
The PXA125 will return the present internal trigger timer value. The returned value will be in standard
scientific format (for example: 10ms would be returned as 10E-3 – positive numbers are unsigned). Default
value is 1e3.
TRIGger:IMMediate
Purpose
This command will stimulate the PXA125 output once. This command will affect the PXA125 only after it will
be set to interrupted operation.
*TRG
Purpose
This common command has the same affect as the TRIG:IMM command.