Teledyne LeCroy PXA125 User Manual

Page 154

Advertising
background image

User Manual PXA125

Programming Reference 4-36

APPLy:RAMP?

Response
The PXA125 will return <freq>, <ampl>,<offs>,<delay>,<high>,<fall>. Default values are

1e6,5,0,10,10,10

APPLy:SINC{<freq>,<ampl>,<offs>,<N_cycles>}

Purpose
This command is a high-level command that programs the PXA125 to output sinc waveform along with its
associated parameters

<freq> sets the output frequency in units of hertz

<ampl> sets the output amplitude in units of volts

<offs> sets the output offset in units of volts

<N_cycles> sets the number of zero crossing

Parameter type

Numeric

Parameter range

<freq>

10e3 to 50e6

<ampl>

80e-3 to 80

<offs>

-3.6 to +3.6

<N_cycle> 4 to 100

APPLy:SINC?

Response
The PXA125 will return <freq>, <ampl>,<offs>,<N_cycle>. Default values are

1e6,5,0,10

APPLy:EXPonential{<freq>,<ampl>,<offs>,<exp>}

Purpose
This command is a high-level command that programs the PXA125 to output exponential waveform along
with its associated parameters

<freq> sets the output frequency in units of hertz

<ampl> sets the output amplitude in units of volts

<offs> sets the output offset in units of volts

<exp> sets the exponent value

Parameter type

Numeric

Parameter range

<freq>

10e3 to 50e6

<ampl>

80e-3 to 80

<offs>

-3.6 to +3.6

<exp>

-20 to +20

APPLy:EXPonential?

Response
The PXA125 will return <freq>, <ampl>,<offs>,<exp>. Default values are

1e6,5,0,1