User Manual PXA125

Programming Reference 4-54

Parameter range

<N_cycle>

4 to 100 zero crossings

SINC:NCYCle?

Response and default
The PXA125 will return the present number of zero-crossing value. Default value is 10.

GAUSsian:EXPonent<exp>

Purpose
This command programs the exponent for the standard gaussian pulse waveform. This command has no
affect on arbitrary waveforms.

<exp > sets the exponent parameter

Parameter type

Numeric

Parameter range

<exp>

10 to 200

GAUSsian:EXPonent?

Response and default
The PXA125 will return the present exponent value. Default value is 20.

EXPonential:EXPonent<exp>

Purpose
This command programs the exponent for the standard exponential waveform. This command has no affect
on arbitrary waveforms.

<exp > sets the exponent parameter

Parameter type

Numeric

Parameter range

<exp>

-100 to 100

EXPonential:EXPonent?

Response and default
The PXA125 will return the present exponent value. Default value is –20.

DC<%_amplitude>

Purpose
This command programs the amplitude of the standard DC waveform. This command has no affect on
arbitrary waveforms.

<%_amplitude > sets the dc level as percentage of the programmed amplitude setting

Parameter type

Numeric