User Manual PXA125

Programming Reference 4-40

FM:FUNCtion:MODE?

Response and default
The PXA125 will return FIX, or USER depending on its present setting. Default is FIX.

FM:FUNCtion:SHAPe{SINusoid|TRIangle|SQUare|RAMP}

Purpose
This command will select one of the waveform shapes as the active modulating waveform.

•

SINusoid will select sine shape as modulating waveform

•

TRIangle will select triangular shape as modulating waveform

•

SQUare will select square shape as modulating waveform

•

RAMP will select ramp shape as modulating waveform

Parameter type

Discrete

FM:FUNCtion:SHAPe?

Response and default
The PXA125 will return SIN, TRI, SQU, or RAMP depending on its present setting. Default is SIN.

FM:FREQuency<FM_freq>

Purpose

This command will set the modulating wave frequency. This parameter affects the PXA125 when set to op-
erate in FM:FUNC:MODE FIX

•

<FM_freq> will set the frequency of the modulating waveform

Parameter type

Numeric

Parameter range

<FM_freq>

1e-3 to 100e6 in units of hertz

FM:FREQuency?

Response and default
The PXA125 will return the present modulating waveform frequency value. The returned value will be in
standard scientific format (for example: 100mHz would be returned as 100E-3 – positive numbers are
unsigned). Default value is 1e3.

FM:FREQuency:RASTer<FM_sclk>

Purpose
This command will set the modulating wave sample clock frequency. This parameter affects the PXA125
when set to operate in FM:FUNC:MODE USER

•

<FM_sclk> will set the sample clock frequency of the modulating waveform

Parameter type

Numeric