IRM-5000P USER’S MANUAL
REV 2
8
2.0 PRE-TEST SETUP
2.1 Operating
Voltages
The IRM-5000P’s operating voltage is preset at the factory and is selectable between 100-120 
Vac, 50/60 Hz or 200-240 Vac, 50/60 Hz. The voltage is set by placing jumpers at the JP16 and J1 
jumper blocks on the main board as listed in Table 3 and illustrated in Figure 2 and Figure 3. 
Table 3. Voltage Selection Jumper Settings
Voltage
Selection
JP16 Settings
J1 Settings
100 – 120 Vac
Jumper Pin 1 and 2 
Jumper Pin 3 and 4 
Jumper Pin 1 and 2 
Jumper Pin 3 and 4 
200 – 240 Vac
Jumper Pin 2 and 3
Jumper Pin 2 and 3
Figure 2. 100-120 Vac Jumper Settings
Figure 3. 200-240 Vac Jumper Settings