Vanguard IRM-5000P User Manual
Page 27
 
REV 2 IRM-5000P USER’S MANUAL
23
NOTE
If you press any other key besides the
[ENTER]
key, you will be
returned to the beginning of step c.
The following screen will be displayed:
Enter the test duration using the numeric keypad and then press the
[ENTER]
key. Continue to step d.
d. The following screen will be displayed:
Select the desired test voltage by pressing the corresponding key (
[1]
,
[2]
,
[3]
,
[4]
, or
[5]
).
e. The following screen will be displayed:
1. YES
Press the
[1]
key (YES) if you would like to enter the Device Under Test (D.U.T.)
temperature. The following screen will be displayed:
Type the D.U.T. temperature using the numeric keypad. The screen will be 
updated as shown: 
Press the
[ENTER]
key and continue to step f.
ENTER TEMP (0-254 F)
 120 
°F
 
TERMINAL VTG: 1 V 
ENTER TEMP (0-254 F)
 
°F
 
TERMINAL VTG: 1 V 
ENTER D.U.T. TEMP?
1. YES 2. NO 
 
TERMINAL VTG: 1 V 
1. 500V 2. 1000V
3. 2500V 4. 5000V 
5. Custom 
TERMINAL VTG: 1 V 
Test Time (Minutes):
 0 Minutes 
 
TERMINAL VTG: 1 V